A2.Age at time of event: (b)(6) years or older.Device evaluated by manufacturer: the device was returned for analysis.A rotawire drive within the returned rotapro device was returned for analysis.The proximal end, middle section, distal end, and spring tip were visually and microscopically examined.Inspection of the device found a kink at 37cm from the distal end of the rotawire.Functional testing was performed using the returned rotawire.During testing, the returned rotawire was able to be removed with resistance, but was not able to be fully reinserted due to the kink in the rotawire.Product analysis confirmed the reported events, as the returned rotawire was able to be removed with resistance but was not able to be reinserted due to the kink in the rotawire.
|