This device is classified as import for export; therefore, 510k is not applicable.Model: eb-1575k is available in the usa with a 510k number: k131028.We checked the returned unit and confirmed that the ccd driver pcb as defective.Based on the result, we concluded that it was caused due to the excessive force applied on the ccd driver pcb.In addition, we confirmed that the electrical connector fluid damage, and the lg cable connector fluid damage; however, they are not the main cause and/or irrelevant to the alleged complaint.Based on the technical report, ""hr-rpt-0586(image failure)"" and/or the risk analysis results, it was evaluated to submit mdr.
|