• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VERATHON MEDICAL ULC GLIDESCOPE CORE SMART CABLE; LARYNGOSCOPE, RIGID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VERATHON MEDICAL ULC GLIDESCOPE CORE SMART CABLE; LARYNGOSCOPE, RIGID Back to Search Results
Model Number 0600-0783
Device Problems Erratic or Intermittent Display (1182); No Display/Image (1183)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/01/2022
Event Type  malfunction  
Event Description
The customer reported that during a patient procedure, using a glidescope core smart cable, the image was reported as inconsistent, and the image would intermittently disappear.The procedure was completed using a backup glidescope core smart cable, which was made readily available.No harm to the patient was reported.
 
Manufacturer Narrative
The device has been received by verathon, however; at the time of the report the device has not been evaluated.Verathon continues to investigate the reported event and a supplemental report will be submitted in accordance with 21 cfr 803.56 when additional information becomes available.
 
Manufacturer Narrative
The reported glidescope core smart cable was returned to verathon for evaluation.A verathon technical service representative evaluated the reported smart cable and was able to confirm the reported image issue.When connecting the reported smart cable to known, good, test verathon equipment, the cable failed to be recognized.Visible corrosion was identified on the cable's hdmi connector pins.The glidescope and gliderite products reprocessing manual states: "use hospital-grade clean air to blow remaining moisture out of the connectors, and then dry the component using hospital-grade clean air." it is likely that not blowing out the connector with hospital-grade air caused or contributed to the corrosion in the hdmi connector.Verathon followed up with the customer and restated the importance of blowing out the connectors following the reprocessing of the cable.Upon completion of the evaluation, the smart cable was scrapped due to having already been replaced prior to its return to verathon and there being no repairs available for this device.Corrective action is not required at this time.Verathon will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLIDESCOPE CORE SMART CABLE
Type of Device
LARYNGOSCOPE, RIGID
Manufacturer (Section D)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, british columbia V5C 5 A9
CA  V5C 5A9
Manufacturer (Section G)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, british columbia V5C 5 A9
CA   V5C 5A9
Manufacturer Contact
corey kasbohm
20001 n creek pkwy
bothell, WA 98011-8218
4256295760
MDR Report Key16051957
MDR Text Key308540769
Report Number9615393-2022-00229
Device Sequence Number1
Product Code CCW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 12/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0600-0783
Device Catalogue Number0800-0602
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/01/2022
Initial Date FDA Received12/27/2022
Supplement Dates Manufacturer Received12/29/2022
Supplement Dates FDA Received01/03/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/21/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-