Model Number 0600-0783 |
Device Problems
Erratic or Intermittent Display (1182); No Display/Image (1183)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/01/2022 |
Event Type
malfunction
|
Event Description
|
The customer reported that during a patient procedure, using a glidescope core smart cable, the image was reported as inconsistent, and the image would intermittently disappear.The procedure was completed using a backup glidescope core smart cable, which was made readily available.No harm to the patient was reported.
|
|
Manufacturer Narrative
|
The device has been received by verathon, however; at the time of the report the device has not been evaluated.Verathon continues to investigate the reported event and a supplemental report will be submitted in accordance with 21 cfr 803.56 when additional information becomes available.
|
|
Manufacturer Narrative
|
The reported glidescope core smart cable was returned to verathon for evaluation.A verathon technical service representative evaluated the reported smart cable and was able to confirm the reported image issue.When connecting the reported smart cable to known, good, test verathon equipment, the cable failed to be recognized.Visible corrosion was identified on the cable's hdmi connector pins.The glidescope and gliderite products reprocessing manual states: "use hospital-grade clean air to blow remaining moisture out of the connectors, and then dry the component using hospital-grade clean air." it is likely that not blowing out the connector with hospital-grade air caused or contributed to the corrosion in the hdmi connector.Verathon followed up with the customer and restated the importance of blowing out the connectors following the reprocessing of the cable.Upon completion of the evaluation, the smart cable was scrapped due to having already been replaced prior to its return to verathon and there being no repairs available for this device.Corrective action is not required at this time.Verathon will continue to monitor for trends.
|
|
Search Alerts/Recalls
|
|
|