Model Number M00550601 |
Device Problem
Display or Visual Feedback Problem (1184)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/29/2022 |
Event Type
malfunction
|
Event Description
|
It was reported to boston scientific corporation that an alliance syringe was used in the common bile duct (cbd) during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2022.During the procedure, the papilla was dilated; however, it was noticed that the pointer of the pressure gauge was inaccurate.The procedure was completed with another alliance syringe.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
|
|
Manufacturer Narrative
|
(b)(4).
|
|
Manufacturer Narrative
|
Block h6: device code a0902 captures the reportable event of the inflation device reading inaccurately.Block h10: investigation results: the returned alliance syringe was analyzed and a visual examination found that the device was returned without its extension tube.There were no visual defects noted on the device.The gauge needle was at 0 atm when received.Microscopic evaluation of the device revealed that the y connector had a fracture close to the connection between the syringe and the y connector.Functional inspection was attempted to pressurize the device at 10 atm with 35 ml of water; however, it was not possible due to a leak in the y connector.No other problems with the device were noted.With all the available information, boston scientific concludes the reported event of gauge reading inaccurate was confirmed as a leak was found and the device cannot hold stable pressure.The leak found in the y connector is likely to have occurred due to the manner in which the device was handled and manipulated.It is possible that excessive manipulation of the device without enough care could have induced the defect found.Therefore, the most probable root cause is adverse event related to procedure.
|
|
Event Description
|
It was reported to boston scientific corporation that an alliance syringe was used in the common bile duct (cbd) during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2022.During the procedure, the papilla was dilated; however, it was noticed that the pointer of the pressure gauge was inaccurate.The procedure was completed with another alliance syringe.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
|
|
Search Alerts/Recalls
|