BOSTON SCIENTIFIC CORPORATION ALLIANCE II; SYRINGE, BALLOON INFLATION Back to Search Results
Model Number M00550601
Device Problem Display or Visual Feedback Problem (1184)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/29/2022
Event Type  malfunction  
Event Description
It was reported to boston scientific corporation that an alliance syringe was used in the common bile duct (cbd) during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2022.During the procedure, the papilla was dilated; however, it was noticed that the pointer of the pressure gauge was inaccurate.The procedure was completed with another alliance syringe.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Manufacturer Narrative
(b)(4).
 
Manufacturer Narrative
Block h6: device code a0902 captures the reportable event of the inflation device reading inaccurately.Block h10: investigation results: the returned alliance syringe was analyzed and a visual examination found that the device was returned without its extension tube.There were no visual defects noted on the device.The gauge needle was at 0 atm when received.Microscopic evaluation of the device revealed that the y connector had a fracture close to the connection between the syringe and the y connector.Functional inspection was attempted to pressurize the device at 10 atm with 35 ml of water; however, it was not possible due to a leak in the y connector.No other problems with the device were noted.With all the available information, boston scientific concludes the reported event of gauge reading inaccurate was confirmed as a leak was found and the device cannot hold stable pressure.The leak found in the y connector is likely to have occurred due to the manner in which the device was handled and manipulated.It is possible that excessive manipulation of the device without enough care could have induced the defect found.Therefore, the most probable root cause is adverse event related to procedure.
 
Event Description
It was reported to boston scientific corporation that an alliance syringe was used in the common bile duct (cbd) during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2022.During the procedure, the papilla was dilated; however, it was noticed that the pointer of the pressure gauge was inaccurate.The procedure was completed with another alliance syringe.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLIANCE II
Type of Device
SYRINGE, BALLOON INFLATION
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
cork business technology park
model farm road
cork
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key16052151
MDR Text Key308204691
Report Number3005099803-2022-07607
Device Sequence Number1
Product Code MAV
UDI-Device Identifier08714729129332
UDI-Public08714729129332
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K922573
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/09/2023
Device Model NumberM00550601
Device Catalogue Number5060-05
Device Lot Number0027635338
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/02/2022
Initial Date FDA Received12/27/2022
Supplement Dates Manufacturer Received01/13/2023
Supplement Dates FDA Received02/03/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/09/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age50 YR
Patient SexMale
Patient Weight63 KG
-
-