Model Number ROB10013 |
Device Problems
Connection Problem (2900); Unintended Application Program Shut Down (4032)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/05/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
Internal complaint reference: (b)(4).
|
|
Event Description
|
It was reported that, during the set-up for a cori assisted tka surgery, the drill attachment would not unlock to place the bur.While unlocking it made a particular sound instead it making a clicking sound.The drill attachment did not lock onto the real intelligence robotic drill.Both devices were swapped and no further issues calibrating were found.During the case, just before cuts were made, they were was kicked out of the case.After the console was restarted a critical error was received and when the drill was switched, it still wouldn¿t work.Surgery was performed manually after a non-significant delay.No harm to the patient or further complications reported.
|
|
Manufacturer Narrative
|
D10, h3, h6: the real intelligence robotic drill, part number rob10013, serial number (b)(6), intended for treatment was not returned for evaluation, therefore a device analysis was unable to be performed.A complaint history review for similar reported/confirmed complaints has identified prior events.A review of manufacturing records indicate the device met all specifications upon release into distribution.Refer to the real intelligence cori for knee arthroplasty user manual, section assembling the robotic drill for proper set up and handling.The failure mode and associated risk have been anticipated within the risk file and the documented risk level is still within the product risk profile.A historical escalation event review was not completed.The product was not returned and no evidence was made available to link the complaint to an escalation event.Although the reported problem was not confirmed through a visual or functional evaluation, factors contributing to the reported symptom may have been associated with an exposure motor encoder failure.Based on the investigation, the need for a corrective action is not recommended or required at this time.Should any additional information be received the complaint will be reopened.The failure mode will continue to be closely monitored through complaint investigation and trended through post market surveillance activities.
|
|
Manufacturer Narrative
|
H3, h6: the cori, real intelligence robotic drill, part number rob10013, (b)(6), used for treatment was returned for evaluation.Nothing was identified visually that contributed to the reported problem.A functional evaluation was performed.The reported problem was not confirmed.The drill passed kpc and a case several times with no errors or connection issues.Disassembly revealed nothing unusual.The exposure motor passed multimeter, portescap, and oscilloscope testing.Although the reported problem was not confirmed through the visual or functional evaluation, no reasonable contributing factors could be identified based on the received complaint information and investigation results.A review of manufacturing records indicates the software met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints found similar events.A historical escalation event review was completed.A review of prior escalation actions found no actions applicable to the scope of the reported complaint.Refer to the real intelligence cori for knee arthroplasty user manual, section assembling the robotic drill for proper set up and handling.The failure mode and associated risk have been anticipated within the risk file and the documented risk level is still adequate.Although no further containment or corrective action is recommended or required at this time, the failure mode will continue to be closely monitored through complaint investigation and trended through post market surveillance activities.
|
|
Manufacturer Narrative
|
H3, h6: the cori, real intelligence robotic drill, part number rob10013, (b)(6), used for treatment was returned for evaluation.Nothing was identified visually that contributed to the reported problem.A functional evaluation was performed.The reported problem was not confirmed.The drill passed kpc and a case several times with no errors or connection issues.Disassembly revealed nothing unusual.The exposure motor passed multimeter, portescap, and oscilloscope testing.Although the reported problem was not confirmed through a visual or functional evaluation, factors that may have contributed to the reported symptom may be associated with an intermittent failure of the drill exposure motor due to the thermo-mechanical stress induced within the motor at the encoder and electrical noise on the console error status inputs to the drill exposure motor encoder.A review of manufacturing records indicates the software met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints found similar events.We have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A historical review concluded that the lot, serial number or part number reported in this event is related to a corrective/preventive action already implemented.Refer to the real intelligence cori for knee arthroplasty user manual, section assembling the robotic drill for proper set up and handling.The failure mode and associated risk have been anticipated within the risk file and the documented risk level is still adequate.Continuous improvements have been made to the cori robotic drill and manufacturing processes to reduce drill disconnection error messages.These improvements consisted of: 1.A hardware update to the cori console to reduce noise on the internal electronics.2.An update to the cori system¿s software and firmware to improve the user experience when error messages are displayed, and 3.A hardware update to the cori drill to reduce mechanical stress on drill exposure the motor.The first two improvements are fully deployed.The third improvement is being deployed for new orders and as drills are returned for routine servicing.Also, smith+nephew is voluntarily performing a recall/field notification for the cori real intelligence robotic drill.Should any additional information be received the complaint will be reopened.The failure mode will continue to be closely monitored through complaint investigation and trended through post market surveillance activities.
|
|