• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BLUE BELT TECHNOLOGIES REAL INTELLIGENCE CORI; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BLUE BELT TECHNOLOGIES REAL INTELLIGENCE CORI; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number ROB10024
Device Problem Computer Operating System Problem (2898)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/05/2022
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event Description
It was reported that a real intelligence cori was not getting booted up.In addition, a real intelligence 24 in.Touch screen did not display any image.Since this was noticed during a preventative maintenance, there was not any patient involved.
 
Manufacturer Narrative
H3, h6.The real intelligence cori, part # rob10024, serial # sn000023, intended for use in treatment, was not returned for evaluation, therefore a visual inspection could not be performed.Visual inspection and functional testing could not be performed since the device was not returned for evaluation.Thus, the complaint could not be verified, nor could a root cause be determined with confidence.Although the reported problem was not confirmed through a visual or functional evaluation, possible contributing factors may have been related to the 24" touchscreen or a power failure in the cori console.A review of manufacturing and service records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints found similar events.A historical escalation event review was not completed.The product was not returned and no evidence was made available to link the complaint to an escalation event.The failure mode and associated risk have been anticipated within the risk file and the documented risk level is still adequate.Although no further containment or corrective action is recommended or required at this time, the failure mode will continue to be closely monitored through complaint investigation and trended through post market surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REAL INTELLIGENCE CORI
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key16055281
MDR Text Key306909768
Report Number3010266064-2022-00688
Device Sequence Number1
Product Code OLO
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K193120
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberROB10024
Device Catalogue NumberROB10024
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/05/2022
Initial Date FDA Received12/27/2022
Supplement Dates Manufacturer Received02/12/2023
Supplement Dates FDA Received02/14/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-