MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND HEATER/COOLER 3T 110V/60 HZ; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND HEATER/COOLER 3T 110V/60 HZ; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-85
Device Problem Temperature Problem (3022)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/30/2022
Event Type  malfunction  
Manufacturer Narrative
Livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in united states.As per field service engineer, patient pumps need to be replaced.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova deutschland received a report that a heater-cooler system 3t displayed error code "pump is blocked or too slow" (error e07) during maintenance.There was no patient involvement.
 
Event Description
See initial report.
 
Manufacturer Narrative
H10: through follow-up communication livanova learned that the error message associated to pump was on patient side of the heater cooler 3t system.A livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported issue.In order to fix it, stirrer motor was replaced.In addition, during repair it was noted that patient pumps were noisy, therefore the were replaced.Subsequent functional verification testing was completed without further issues and the unit was returned to service.Complaints database analysis revealed that no similar event on this device occurred since its installation in 2015.The symptoms (noisy and stuck motors) found by the livanova technician on the replaced components are related to a mechanical fault.The most likely root cause of the reported issue is associated to motor bearing damaged by the corrosion, which consequently does not allow the motor shaft to rotate.Possible causes are water infiltration or water formation due to condensation or high temperatures and high level of humidity in the stationary phase.Causes related to environmental conditions.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER/COOLER 3T 110V/60 HZ
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80939
GM   80939
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key16059172
MDR Text Key307578477
Report Number9611109-2022-00691
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K191402
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup
Report Date 11/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-85
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/30/2022
Initial Date FDA Received12/28/2022
Supplement Dates Manufacturer Received01/09/2023
Supplement Dates FDA Received02/08/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/27/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-