Patient identifier,age or date of birth,sex,weight,ethnicity: there was no known patient involvement.The heater-cooler 16-02-80 is not distributed in the usa and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402).H10: livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in indonesia as per follow up with the customer, initially the heater-cooler was noisy and showed the error 06 on the warm and cold cardioplegia pump.The hospital field engineer substituted the circulator motor em109, distribution board and cpu board.The noise issue was solved.The error 06 on the warm cardioplegia displays was not solved.Additionally, after the service intervention for new parts, the error 17 was shown on the display of the patient tank.The field engineer has use esd protection when installed the circulation board and cpu board.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|