Model Number 16-02-80
Device Problem Temperature Problem (3022)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/25/2022
Event Type  malfunction  
Manufacturer Narrative
Patient identifier,age or date of birth,sex,weight,ethnicity: there was no known patient involvement.The heater-cooler 16-02-80 is not distributed in the usa and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402).H10: livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in indonesia as per follow up with the customer, initially the heater-cooler was noisy and showed the error 06 on the warm and cold cardioplegia pump.The hospital field engineer substituted the circulator motor em109, distribution board and cpu board.The noise issue was solved.The error 06 on the warm cardioplegia displays was not solved.Additionally, after the service intervention for new parts, the error 17 was shown on the display of the patient tank.The field engineer has use esd protection when installed the circulation board and cpu board.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova deutschland received a report that a heater-coolersystem 3t had the distribution board ul510 damaged when installed.Initially, the error e6 on the warm cardioplegia circuit was displayed and then error e17 on the patient circuit was also displayed.There was no patient involvement.
 
Event Description
See initial report.
 
Manufacturer Narrative
A complaints database review has been performed and no further similar issue has been submitted in the past, therefore the reported put of box failure is an isolated event and any systematic quality issues can be excluded.The root cause of e17 error code has been traced back to a faulty distribution board.The risk is in the acceptable region.No specific action was currently deemed necessary.Livanova will keep monitoring the market.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w 65th way
arvada, CO 80004
MDR Report Key16059451
MDR Text Key308242403
Report Number9611109-2022-00693
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeID
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup
Report Date 03/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-80
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/01/2022
Initial Date FDA Received12/28/2022
Supplement Dates Manufacturer Received02/11/2023
Supplement Dates FDA Received03/10/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/26/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-