• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. VISERA CYSTO-NEPHRO VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. VISERA CYSTO-NEPHRO VIDEOSCOPE Back to Search Results
Model Number CYF-V2
Device Problem Device Reprocessing Problem (1091)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/07/2022
Event Type  malfunction  
Manufacturer Narrative
The device was returned and evaluated and the customers allegation was confirmed.Additionally, the angulation rubber was leaking, the cable tube unit was wrinkling, water tightness was lost due to cut on the bending section cover, and the adhesive on the bending section cover was detached.It was recommended that the bending rubber and cable tube both be replaced.The investigation is ongoing and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event Description
The customer reported to olympus that while reprocessing the visera cysto-nephro videoscope, the scope had mistakenly gone into the sterilizer without the eto cap; as a result, the insulation and fluid had gotten inside the scope.There was no procedural involvement or patient harm associated with the event.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the definitive root cause of the improper scope placement during reprocessing could not be determined.The instruction manual identifies the following verbiage, which may have prevented the phenomenon: ¿7.11 sterilization ethylene oxide gas sterilization [warning] ·attach the eto cap to the venting connector before sterilizing.If the eto cap is not attached to the endoscope during sterilization, the vacuum created within the sterilization chamber can rupture the covering of the bending section.¿ olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISERA CYSTO-NEPHRO VIDEOSCOPE
Type of Device
CYSTO-NEPHRO VIDEOSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16059902
MDR Text Key306740590
Report Number3002808148-2022-05659
Device Sequence Number1
Product Code FAJ
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K133538
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCYF-V2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/11/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/07/2022
Initial Date FDA Received12/28/2022
Supplement Dates Manufacturer Received01/23/2023
Supplement Dates FDA Received02/10/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/12/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-