• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® ALL-SILICONE TEMPERATURE-SENSING FOLEY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® ALL-SILICONE TEMPERATURE-SENSING FOLEY CATHETER Back to Search Results
Model Number 119112
Device Problem Difficult to Remove (1528)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/29/2022
Event Type  malfunction  
Event Description
It was reported that the water could be drained only halfway when the user tried to remove the water from the foley catheter during the pretest.About 2 ml of water was aspirated by syringe.It was stated that the user tried to deflate using syringe but the balloon was partially deflated and the catheter could be removed.Completely.Usually, balloon can be deflated within a few seconds.
 
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Manufacturer Narrative
The reported event was confirmed and the cause was unknown.The potential failure modes are collapsed lumen and blocked lumen.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "[warnings].1.Method for use.(1) do not inflate the balloon in the urethra.[the urethra may be injured.] (2) do not pull the catheter hard.[the bladder/urethra may be injured.] 2.Applicable patients.Patients with delirium who might pull out catheter [when patient tugs at catheter unconsciously, the bladder and urethra may be damaged.] [contraindications].1.Method for use: (1) do not reuse.(2) do not resterilize.(3) this device contains 10% povidone-iodine.For patients with past history of allergic hypersensitivity to povidone-iodine or iodine, consider using alternative disinfectants.(4) be careful that the catheter is not exposed to ointments, contrast medium or oil-based lubricants (including vegetable oils such as olive oil, mineral oils such as white petrolatum and animal oils).[they may damage the device and may burst balloon.] (5) do not hold the device with forceps, etc.Avoid contact with any blades or sharp-edged instruments.[catheter damage may cause balloon rupture and accidental balloon removal or failure to deflate or remove the balloon.] 2.Applicable patients.Patients with known allergy to silver coated catheter".H11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.H3 other text : the actual/suspected device was inspected.
 
Event Description
It was reported that the water could be drained only halfway when the user tried to remove the water from the foley catheter during the pretest.About 2 ml of water was aspirated by syringe.It was stated that the user tried to deflate using syringe but the balloon was partially deflated and the catheter could be removed completely.Usually, balloon can be deflated within a few seconds.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® ALL-SILICONE TEMPERATURE-SENSING FOLEY CATHETER
Type of Device
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key16060966
MDR Text Key308471053
Report Number1018233-2022-09902
Device Sequence Number1
Product Code EZL
UDI-Device Identifier00801741046018
UDI-Public(01)00801741046018
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K070582
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,User Facility
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number119112
Device Catalogue Number119112
Device Lot NumberNGGP1733
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/21/2022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/12/2022
Initial Date FDA Received12/28/2022
Supplement Dates Manufacturer Received03/15/2023
Supplement Dates FDA Received03/22/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/21/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-