It was reported to boston scientific corporation on (b)(6) 2022 that an ultraflex esophageal proximal covered stent was to be implanted to treat an 8cm malignant stricture in the middle esophagus during a stent placement procedure performed on (b)(6) 2022.The patient's anatomy was dilated prior to stent placement.During the procedure, the stent was deployed successfully; however, it was noted that the stent was unable to expand.Consequently, a cre balloon dilatation was used to facilitate stent expansion but the result was not good.An upper gi endoscopy was performed after 48hrs and it was noted that the stent was able to expand partially with kink in the middle.Subsequently, the stent was removed with a rat-tooth foreign body forceps and a different device was used to complete the procedure.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.Note: photos of the explanted ultraflex esophageal stent were provided and it was observed that the stent was unable to expand completely and was damaged.
|