SHIRAKAWA OLYMPUS CO., LTD. EVIS EXERA III VIDEO SYSTEM CENTER; ENDOSCOPIC VIDEO IMAGE PROCESSING UNIT Back to Search Results
Model Number CV-190
Device Problem No Display/Image (1183)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/01/2022
Event Type  malfunction  
Event Description
The customer reported that scopes fail to work with the evis exera iii video system center.The customer attempted a different maj-1430 however the issue persisted.The event occurred during preparation for use and there was no patient harm or user injury reported due to the event.
 
Manufacturer Narrative
Olympus technical assistance center (tac) support spoke to the customer to troubleshoot the device.Tac concluded that the cv-190 socket is bad and advised the customer to blow out dust before attempting a 190 series scope.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time, however, if additional information becomes available, this report will be supplemented accordingly.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 6 years since the subject device was manufactured.Based on the results of the legal manufacturer's investigation, and the inability to evaluate the subject device and reproduce the event, a definitive root cause of the reported event could not be identified.However, it¿s likely the cause is related to a failure to receive the video jacket of the device.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III VIDEO SYSTEM CENTER
Type of Device
ENDOSCOPIC VIDEO IMAGE PROCESSING UNIT
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16064255
MDR Text Key306637293
Report Number3002808148-2022-05697
Device Sequence Number1
Product Code NWB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131780
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 03/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCV-190
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/01/2022
Initial Date FDA Received12/28/2022
Supplement Dates Manufacturer Received02/08/2023
Supplement Dates FDA Received03/03/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/24/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
MAJ-1430
-
-