• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OGDEN MANUFACTURING PLANT OPTIFLUX 160NRE DIALYZER FINISHED ASSY.; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OGDEN MANUFACTURING PLANT OPTIFLUX 160NRE DIALYZER FINISHED ASSY.; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Model Number 0500316E
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/14/2022
Event Type  malfunction  
Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.
 
Event Description
A registered nurse (rn) reported that a dialyzer blood leak occurred during a hemodialysis (hd) treatment.In a follow up with registered nurse (rn) it was clarified that the blood leak occurred right at the start of the hd treatment.The blood leak was described as being an internal blood leak.The leak was visually observed in the dialysate lines/hansens.The machine, a fresenius 2008t machine, alarmed appropriately with a blood leak alarm.Blood leak test strips were not used.Fresenius bloodlines were used.There was no damage noted on the dialyzer.The patient¿s estimated blood loss (ebl) was < 300 ml.There was no patient injury, adverse events, or medical intervention required as a result of this event.The treatment was completed on a different machine with new supplies.The device is available to be returned to the manufacturer for evaluation.
 
Manufacturer Narrative
Plant investigation: a sample has not been provided for evaluation.The third party carrier's website was reviewed and it was verified that the provided shipping materials were sent to the customer and were subsequently received.In the event a sample is returned for evaluation, the complaint file will be updated the reported complaint was not confirmed as the complaint device was not returned for manufacturer evaluation.A production records review was performed on the reported lot.An investigation of the device history records (dhr) was conducted by the manufacturer.There was one approved temporary deviation notice (dn) reported on the lot which was unrelated to the complaint event.There was no indication of product nonacceptance, deviation, non-conformance, rework, labeling or process control failure during the manufacturing process which could be associated with the reported event.The lot met all release criteria.A definitive conclusion regarding the complaint incident cannot be reached without physical examination of the actual device.Therefore, the complaint is not confirmed.Continuous improvement is of the utmost importance to fresenius medical care as we strive to provide dialysis products of the highest quality to our patients.Reports of leaking product are investigated both individually as complaints, as well as via the nc/capa program in order to assess and improve our products and processes.Capas for vision systems and blood leak reduction are recent examples of leak related investigations directed at an overall reduction in dialyzer leaks.
 
Event Description
A registered nurse (rn) reported that a dialyzer blood leak occurred during a hemodialysis (hd) treatment.In a follow up with registered nurse (rn) it was clarified that the blood leak occurred right at the start of the hd treatment.The blood leak was described as being an internal blood leak.The leak was visually observed in the dialysate lines/hansens.The machine, a fresenius 2008t machine, alarmed appropriately with a blood leak alarm.Blood leak test strips were not used.Fresenius bloodlines were used.There was no damage noted on the dialyzer.The patient¿s estimated blood loss (ebl) was < 300 ml.There was no patient injury, adverse events, or medical intervention required as a result of this event.The treatment was completed on a different machine with new supplies.The device is available to be returned to the manufacturer for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTIFLUX 160NRE DIALYZER FINISHED ASSY.
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
OGDEN MANUFACTURING PLANT
director, site quality
475 west 13th street
ogden UT 84404
Manufacturer (Section G)
OGDEN MANUFACTURING PLANT
director, site quality
475 west 13th street
ogden UT 84404
Manufacturer Contact
jessica trujillo
920 winter st
waltham, MA 02451
6174175172
MDR Report Key16064356
MDR Text Key307882258
Report Number0001713747-2022-00640
Device Sequence Number1
Product Code KDI
UDI-Device Identifier00840861100149
UDI-Public00840861100149
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162488
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 01/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0500316E
Device Catalogue Number0500316E
Device Lot Number22KU06013
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device AgeMO
Initial Date Manufacturer Received 12/14/2022
Initial Date FDA Received12/28/2022
Supplement Dates Manufacturer Received01/19/2023
Supplement Dates FDA Received01/30/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/19/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
FRESENIUS 2008T MACHINE; FRESENIUS 2008T MACHINE; FRESENIUS BLOODLINES; FRESENIUS BLOODLINES
Patient Age92 YR
Patient SexFemale
Patient Weight64 KG
-
-