Model Number CYF-5 |
Device Problem
Microbial Contamination of Device (2303)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/13/2022 |
Event Type
malfunction
|
Event Description
|
The customer reported to olympus, during reprocessing, the oes cystonephrofiberscope tested positive for 1 colony forming units (cfus) of lysinibacillus fusiformis on (b)(6).On (b)(6), it tested positive for 1 cfu of penicillium chrysogenum.All channels were sampled.The user did not report any contamination or any other serious deterioration in state of health of any person, to which the scope could have been a contributory cause.
|
|
Manufacturer Narrative
|
The suspect device has not yet been returned to olympus for evaluation and the investigation is ongoing.The physical device evaluation has not been completed.Prior to the device evaluation, the device was sent out for additional microbiological testing.The microbiological analysis report indicated the channels of the scope were cultured and found 1 cfu of micrococcaceae.The results obtained did not comply with the target level for an endoscope subjected to high level disinfection and rinsed with sterile water.Once the investigation has been completed, a supplemental report will be submitted with device evaluation results.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the device evaluation and the legal manufacturer's final investigation.The device was evaluated where no abnormalities were found that could have led to the positive culture.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a root cause could not be determined.Growth of microorganisms were found through culture testing by the user after reprocessing.However, when olympus culture tested after reprocessing in accordance with instructions for use (ifu) before repair, the results conformed to the regulation's recommendation.The instruction manual of cyf-5, describes the reprocessing methods in the following chapters: chapter "compatible reprocessing methods and chemical agents".Chapter "cleaning, disinfection, and sterilization procedures".Olympus will continue to monitor field performance for this device.
|
|
Search Alerts/Recalls
|
|
|