Model Number CDS0706-NTW |
Device Problems
Difficult or Delayed Positioning (1157); Entrapment of Device (1212); Incomplete Coaptation (2507)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/08/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
|
Event Description
|
This is filed to report the clip becoming stuck in chordae and a single leaflet device attachment.It was reported that a patient presented with grade 4 functional mitral regurgitation (mr).During a mitraclip procedure, the second clip became stuck and was unable to move in the ventricle.Leaflet capture was not optimal during positioning for deployment.The mr was reduced to grade 2-3.During post-procedure assessment, the anterior leaflet was detached.A single leaflet device attachment (slda) occurred.There were no adverse patient sequelae or clinically significant delay.No additional information was received.
|
|
Manufacturer Narrative
|
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other complaints reported from this lot.The investigation determined the reported difficulty grasping, difficulty capturing and subsequent slda appear to be related to the reported entrapment of device (clip stuck on ventricle).The entrapment of device appears to be related to user technique.The reported unexpected medical intervention was a result of case-specific circumstance.There is no indication of a product issue with respect to manufacture, design or labeling.
|
|
Search Alerts/Recalls
|