• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NIHON KOHDEN CORPORATION BSM-3532A; VITAL SIGNS MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NIHON KOHDEN CORPORATION BSM-3532A; VITAL SIGNS MONITOR Back to Search Results
Model Number BSM-3532A
Device Problems Mechanical Problem (1384); Detachment of Device or Device Component (2907); Device Fell (4014)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/02/2022
Event Type  malfunction  
Manufacturer Narrative
The customer reported that the bedside monitor (bsm) fell from the wall mount it was attached to.The mount was not correctly mounted into the wall studs, and the device fell from about 5 feet up.No harm or injury was reported.They would like to send the unit in to fix the physical damage.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.Additional device information: concomitant medical device: the following device was used in conjunction with the bsm and is the device that experienced the failure: wall support bracket: model #: the customer could not provide this information.Serial #: the customer could not provide this information.
 
Event Description
The customer reported that the bedside monitor (bsm) fell from the wall mount it was attached to.The mount was not correctly mounted into the wall studs, and the device fell from about 5 feet up.No harm or injury was reported.
 
Event Description
The customer reported that the bedside monitor (bsm) fell from the wall mount it was attached to.The mount was not correctly mounted into the wall studs, and the device fell from about 5 feet up.No harm or injury was reported.
 
Manufacturer Narrative
Details of complaint: the customer reported that the bedside monitor (bsm) fell from the wall mount it was attached to.The mount was not correctly mounted into the wall studs, and the device fell from about 5 feet up.No patient harm or injury was reported.Investigation summary: the customer stated that their maintenance personnel installed the mount and that the mount was not correctly mounted into the studs of the wall.No injuries were reported.Based on the available information, the most probable cause of the issue is user error - incorrect installation.Should the mount have been properly installed, the issue may have been prevented.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BSM-3532A
Type of Device
VITAL SIGNS MONITOR
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
attn: shama mooman
1-31-4 nishiochia
shinjuku-ku, tokyo 161-8 560
JA  161-8560
Manufacturer (Section G)
NIHON KOHDEN TOMIOKA CORPORATION
attn: shama mooman
1-1 tajino
tomioka city, gunma 370-2 314
JA   370-2314
Manufacturer Contact
shama mooman
seibu bldg 2, 4th floor 1-11-2
seibu bldg 2, 4th floor 1-11-2
kusunokidai tokorozawa, saitama 359-8-580
JA   359-8580
MDR Report Key16065194
MDR Text Key308349230
Report Number8030229-2022-03265
Device Sequence Number1
Product Code MHX
UDI-Device Identifier04931921113660
UDI-Public04931921113660
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K080342
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 04/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBSM-3532A
Device Catalogue NumberBSM-3532A
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/02/2022
Initial Date FDA Received12/28/2022
Supplement Dates Manufacturer Received03/09/2023
Supplement Dates FDA Received04/18/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/14/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
WALL SUPPORT BRACKET; WALL SUPPORT BRACKET
-
-