MAUDE Adverse Event Report: BLUE BELT TECHNOLOGIES REAL INTELLIGENCE CORI; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BLUE BELT TECHNOLOGIES REAL INTELLIGENCE CORI; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number ROB10024
Device Problem Unintended Application Program Shut Down (4032)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/07/2022
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event Description
It was reported that, during the set-up for a cori assisted surgery, a "critical error: the system has detected that launcher exited due to a configuration error.Please contact robotics customer support" was received.After rebooting twice, the same error appeared.When shutting it down, the "blue man" was noticed on the console.Surgery was performed after a non-significant delay with the same device.Since this was noticed during the set-up for surgery, the patient was not yet involved.
 
Manufacturer Narrative
H3, h6: the real intelligence cori, part number rob10024, serial number (b)(6), used for treatment was not returned for evaluation, therefore a device analysis was unable to be performed.A complaint history review for similar reported/confirmed complaints has identified prior events.A review of manufacturing records indicate the device met all specifications upon release into distribution.The failure mode and associated risk have been anticipated within the risk file and the documented risk level is still within the product risk profile.A historical escalation event review was not completed.The product was not returned and no evidence was made available to link the complaint to an escalation event.Although the reported problem was not confirmed through a visual or functional evaluation, factors contributing to the reported symptom may have been associated with console software bug.Based on the investigation, the need for a corrective action is not recommended or required at this time.Should any additional information be received the complaint will be reopened.The failure mode will continue to be closely monitored through complaint investigation and trended through post market surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REAL INTELLIGENCE CORI
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key16067061
MDR Text Key306909933
Report Number3010266064-2022-00690
Device Sequence Number1
Product Code OLO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K193120
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberROB10024
Device Catalogue NumberROB10024
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/07/2022
Initial Date FDA Received12/29/2022
Supplement Dates Manufacturer Received01/29/2023
Supplement Dates FDA Received01/30/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-