• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 DXTEND METAGLENE; DELTA XTEND IMPLANTS : SHOULDER METAGLENE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 DXTEND METAGLENE; DELTA XTEND IMPLANTS : SHOULDER METAGLENE Back to Search Results
Model Number 130760000
Device Problems Device Dislodged or Dislocated (2923); Osseointegration Problem (3003)
Patient Problem Inadequate Osseointegration (2646)
Event Date 12/13/2022
Event Type  Injury  
Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Reason for revision geosphere had dislocated from metaglene superiorly.Pe cup had worn.Metaglene and screws had become loose.On removal of geosphere the metaglene and screws also came out of the glenoid.Screw sizes unknown as notes not available.All screws removed from patient.3 fell out of the glenoid with the metaglene baseplate.One had snapped but the surgeon removed the snapped element so none was left in the patient.Delta cta head implanted.The humeral stem was left untouched.No notes available from primary operation.Hospital retained implants so nothing to send back.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DXTEND METAGLENE
Type of Device
DELTA XTEND IMPLANTS : SHOULDER METAGLENE
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY INT'L LTD. 8010379
st anthonys road
leeds LS11 8 DT
UK   LS11 8 DT
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key16067915
MDR Text Key306334048
Report Number1818910-2022-26246
Device Sequence Number1
Product Code HSD
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K192855
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number130760000
Device Catalogue Number130760000
Device Lot Number5320933
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/13/2022
Initial Date FDA Received12/29/2022
Supplement Dates Manufacturer Received01/13/2023
Supplement Dates FDA Received01/15/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/01/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DXTEND GLENOSPHERE STD D42MM; DXTEND STAND PE CUP D42 +6MM; UNKNOWN SHOULDER LOCKING SCREW; UNKNOWN SHOULDER LOCKING SCREW; UNKNOWN SHOULDER LOCKING SCREW; UNKNOWN SHOULDER NON-LOCKING SCREW
Patient Outcome(s) Required Intervention;
Patient Age64 YR
Patient SexMale
-
-