Model Number CV-180
Device Problem Erratic or Intermittent Display (1182)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
A user facility reported to olympus that a vertical and horizontal line was noted with the evis exera ii video system center when connecting the gif-h180.During a standard service inspection of the customer returned device, image disappearance occurred.This report is to capture the reportable malfunction found at inspection.There was no patient harm or user injury reported due to the event.
 
Manufacturer Narrative
The device was returned for investigation.Upon evaluation of the device, image disappeared due to a faulty power supply unit.In addition, y/c output connector deformed, receptacle pin deformed, main/front panel chassis corrosion, tracks of cables found rusty, distribution panel board rust, and picture in picture corrosion were observed.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time, however, if additional information becomes available, this report will be supplemented accordingly.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is probable the image issue was caused by a faulty power supply.However, the root cause of the event could not be determined.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II VIDEO SYSTEM CENTER
Type of Device
VIDEO SYSTEM CENTER
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16068324
MDR Text Key308375301
Report Number3002808148-2022-05723
Device Sequence Number1
Product Code FAJ
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K133538
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 02/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCV-180
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/04/2022
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/04/2022
Initial Date FDA Received12/29/2022
Supplement Dates Manufacturer Received02/02/2023
Supplement Dates FDA Received02/27/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/29/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
GIF-H180
-
-