Catalog Number 2C9961KP
Device Problems Melted (1385); Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/25/2022
Event Type  malfunction  
Event Description
It was reported that the administration tubing of a large volume multirate infusor was damaged.The damaged tubing was twisted/bent and described as ¿burnt¿.The issue was identified before use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
H4: the lot was manufactured from july 28, 2022 - june 29, 2022.H10: the actual device was not available; however, a photograph of the sample was provided for evaluation.Visual inspection was performed to the photograph showed a segment of the tubing line has been damaged.Additionally, indentation marks from the flow wrap sealer equipment were observed on the damaged area of the tubing line was observed.The reported condition was verified.The cause of the condition was due to manufacturing related issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue
building 3
irvine CA 92614
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key16068669
MDR Text Key308257623
Report Number1416980-2022-07142
Device Sequence Number1
Product Code MEB
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
K011317
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Consumer,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2C9961KP
Device Lot Number22G034
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/05/2022
Initial Date FDA Received12/29/2022
Supplement Dates Manufacturer Received01/26/2023
Supplement Dates FDA Received02/03/2023
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
NA.
-
-