• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL MITRACLIP G4 CLIP DELIVERY SYSTEM; MITRAL VALVE REPAIR DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL MITRACLIP G4 CLIP DELIVERY SYSTEM; MITRAL VALVE REPAIR DEVICES Back to Search Results
Model Number CDS0701-XTW
Device Problem Unintended Movement (3026)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/07/2022
Event Type  malfunction  
Event Description
This report is being filed as the clip opened while locked.It was reported that on (b)(6) 2022, during preparation of a mitraclip, the mitraclip arms would open while establishing final arm angle.The lock lever had been down, and a full rotation of the arm positioner after neutral was performed.The clip however, would open while establishing final arm angle.There was no patient involvement, and a new device was prepped and used.No additional information was provided.
 
Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar incidents reported from this lot.Based on available information and without the device to analyze, the cause of the reported unintended movement (clip open ¿ efaa) could not be determined.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP G4 CLIP DELIVERY SYSTEM
Type of Device
MITRAL VALVE REPAIR DEVICES
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key16069524
MDR Text Key308500934
Report Number2135147-2022-02755
Device Sequence Number1
Product Code NKM
UDI-Device Identifier08717648231001
UDI-Public08717648231001
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/07/2023
Device Model NumberCDS0701-XTW
Device Catalogue NumberCDS0701-XTW
Device Lot Number20908R1051
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/07/2022
Initial Date FDA Received12/29/2022
Supplement Dates Manufacturer Received01/10/2023
Supplement Dates FDA Received01/30/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/08/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-