BAXTER HEALTHCARE CORPORATION ULTRAFILTER HECHINGEN; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 114604
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/04/2022
Event Type  malfunction  
Event Description
It was reported that crystallization was observed on two (2) units of u9000 before patient use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
(b)(6).The device was not received for evaluation; however, pictures were received.Visual inspection observed white deposits in the area of the end cap.The ultrafilter is not responsible for the failure of 'particulate matter inside fluid path¿ it only fulfills its function as a water filter.No product defect was observed.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAFILTER HECHINGEN
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - HECHINGEN BADEN DIALYZER
holger - crafoord - strasse 26
hechingen D-723 79
GM   D-72379
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key16070983
MDR Text Key307502401
Report Number9611369-2022-00268
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse Practitioner
Type of Report Initial
Report Date 12/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2023
Device Catalogue Number114604
Device Lot Number2-1801-H-01
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/05/2022
Initial Date FDA Received12/29/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
AK98 (SERIAL NUMBER: UNKNOWN).
-
-