Block b3: approximated based on the date the manufacturer became aware of the event.Block h6: device problem code a0902 captures the reportable event of gauge reading inaccurate.Block h10: investigation results.The returned alliance inflation syringe was analyzed, and a visual evaluation found that there were no damages to the syringe but, the extension tube was blocked.Microscopic inspection confirmed the device was blocked.A functional inspection was performed and the syringe was inflated with water.However, the device did not fill with water due to a blockage in the extension tube on the female side.The device was filled with water without the extension tube and the syringe was pressurized to 10 atm for 30 seconds.The pressure reading on the gauge was accurate and no leaks were found.No other problems with the device were noted.With all the available information, boston scientific concludes the reported event of gauge reading inaccurately was not confirmed.No damages were found on the returned device and the pressure reading was accurate during functional testing without the extension tube.However, a blockage was found in the extension tube of the alliance syringe.This is likely to have occurred due to factors encountered during the manufacturing of the device.Therefore, the most probable root cause is manufacturing deficiency.An investigation to address this problem has been completed.
|