Model Number 106
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Convulsion, Clonic (2222)
Event Date 09/12/2018
Event Type  Injury  
Manufacturer Narrative
Livanova usa, inc.Submits this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation, based on information that livanova has obtained, but may not have been able to investigate or verify prior to the date the report was required by the fda.This report does not constitute an admission, or a conclusion by fda or anyone else, that the device, livanova, or livanova's employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects¿ or "malfunctions¿.These words are incorporated into the fda 3500a medwatch form by the fda, and livanova objects to their use.
 
Event Description
It was reported from clinic notes in 2018 that the patient was noted to have absent seizures during period of illness and gtcs on (b)(6) 2018.The vns settings were adjusted and the patient experienced no further seizures.No other relevant information has been received to date.
 
Event Description
The generator was noted to be replaced prophylactically.The explanted device has not been received to date.No other relevant information has been received to date.
 
Event Description
It was reported that the explanted generator was received by the manufacturer.No other relevant information has been received to date.
 
Event Description
The product had undergone product analysis.There were no performance, or any other type of adverse conditions found with the pulse generator.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PULSE GEN MODEL 106
Type of Device
GENERATOR
Manufacturer (Section D)
LIVANOVA USA, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
LIVANOVA USA, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
dana sprague
100 cyberonics blvd
suite 600
houston, TX 77058
2816672681
MDR Report Key16071089
MDR Text Key306363774
Report Number1644487-2022-01685
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date 03/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date08/31/2018
Device Model Number106
Device Lot Number5011
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received 12/08/2022
Initial Date FDA Received12/29/2022
Supplement Dates Manufacturer Received01/05/2023
02/08/2023
03/10/2023
Supplement Dates FDA Received02/02/2023
03/03/2023
03/22/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/20/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
Patient Age5 YR
Patient SexFemale
-
-