• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC NEUROMODULATION VERCISE GENUS; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC NEUROMODULATION VERCISE GENUS; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS Back to Search Results
Model Number DB-1216
Device Problem Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problem Undesired Nerve Stimulation (1980)
Event Date 12/07/2022
Event Type  Injury  
Event Description
It was reported that the patient experienced shocking sensations when the deep brain stimulator (dbs) implantable pulse generator (ipg) was being charged.The issue did not resolve after trying another ipg charger.Therefore, the patient was administered xanax medication to help make charging the ipg easier.The physician assessed the event may be behavioral, however, the cause is unknown.The device remains implanted in the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERCISE GENUS
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
erik sherburne
25155 rye canyon loop
valencia, CA 91355
7632920920
MDR Report Key16074847
MDR Text Key306404312
Report Number3006630150-2022-07388
Device Sequence Number1
Product Code NHL
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date05/18/2024
Device Model NumberDB-1216
Device Catalogue NumberDB-1216
Device Lot Number541601
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/07/2022
Initial Date FDA Received12/29/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/19/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age66 YR
Patient SexMale
-
-