BECTON DICKINSON MEDICAL (SINGAPORE) BD PEGASUS¿ SAFETY CLOSED IV CATHETER SYSTEM; INTRAVASCULAR CATHETER Back to Search Results
Catalog Number 383714
Device Problem Degraded (1153)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/15/2022
Event Type  malfunction  
Manufacturer Narrative
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported that the bd pegasus¿ safety closed iv catheter system was rusty and the packaging was intact.The following information was provided by the initial reporter, translated from chinese to english: when the nurse was about to give the child an injection, she took out the indwelling needle and found that the "anti-needle stick" device was rusty and the packaging was intact.
 
Manufacturer Narrative
H6: investigation summary: a device history review was conducted for lot number 1141277.Our records show that this is the only instance of this issue occurring in this production batch.According to the sampling plan applied for product performance, this lot was accepted and released without defects being noted during the final assembly or visual inspections.A sample could not be obtained for evaluation and testing; in lieu of the affected device, visual analysis was performed on retention samples for this lot, the retained devices were found to be free of any damage or other abnormalities.Unfortunately without the ability to investigate the affected unit our quality engineers were unable to determine the root cause for this complaint.
 
Event Description
It was reported that the bd pegasus¿ safety closed iv catheter system was rusty and the packaging was intact.The following information was provided by the initial reporter, translated from chinese to english: when the nurse was about to give the child an injection, she took out the indwelling needle and found that the "anti-needle stick" device was rusty and the packaging was intact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD PEGASUS¿ SAFETY CLOSED IV CATHETER SYSTEM
Type of Device
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BECTON DICKINSON MEDICAL (SINGAPORE)
30 tuas avenue 2
singapore
Manufacturer (Section G)
BECTON DICKINSON MEDICAL (SINGAPORE)
30 tuas avenue 2
singapore
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key16075121
MDR Text Key308528405
Report Number8041187-2022-00776
Device Sequence Number1
Product Code FOZ
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date05/31/2024
Device Catalogue Number383714
Device Lot Number1141277
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/08/2022
Initial Date FDA Received12/29/2022
Supplement Dates Manufacturer Received01/17/2023
Supplement Dates FDA Received02/01/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/21/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-