• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND HEATER/COOLER 3T 110V/60 HZ; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND HEATER/COOLER 3T 110V/60 HZ; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-85
Device Problem Temperature Problem (3022)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/02/2022
Event Type  malfunction  
Manufacturer Narrative
Patient information was not provided.Livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in united states.A livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported issue: errore 6 at patient side and circulator motor was not functioning.The circulator motor em109, erosion kit and processor board were replaced.Subsequent functional verification testing was completed without further issues and the unit was returned to service.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova deutschland received a report that a heater-cooler system 3t displayed an error message (error 6) associated to the stirrer motor of patient pump during a procedure.There is not report of any patient injury.
 
Event Description
See initial report.
 
Manufacturer Narrative
B.5.The event occurred while cleaning and disinfecting the device and it did not occured during procedure as previously communicated.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Manufacturer Narrative
H10: most likely root cause of the reported event could be traced back to motor bearing damaged by the corrosion, which consequently does not allow the motor shaft to rotate freely.Possible causes are water infiltration or water formation due to condensation or high temperatures and high level of humidity in the stationary phase.Causes related to environmental conditions.The malfunction of the stirring motor led to damage also the processor board.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER/COOLER 3T 110V/60 HZ
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80939
GM   80939
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key16076517
MDR Text Key308492753
Report Number9611109-2022-00696
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K191402
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup,Followup
Report Date 01/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-85
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/02/2022
Initial Date FDA Received12/30/2022
Supplement Dates Manufacturer Received12/12/2022
03/31/2023
Supplement Dates FDA Received01/11/2023
04/18/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/30/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-