• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOSENSE WEBSTER INC CARTO 3 SYSTEM; COMPUTER, DIAGNOSTIC, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOSENSE WEBSTER INC CARTO 3 SYSTEM; COMPUTER, DIAGNOSTIC, PROGRAMMABLE Back to Search Results
Device Problem Defective Device (2588)
Patient Problem Insufficient Information (4580)
Event Date 12/12/2022
Event Type  malfunction  
Event Description
The carto 3 system malfunctioned.Despite troubleshooting, the room and system were unable to be used.It was later identified by tech support to be a magnet malfunction.The magnet is crucial for mapping and without this the system couldn't be uploaded.The part was ordered overnight, and tech support was on site for installation.Unfortunately, i don't have the device serial number.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARTO 3 SYSTEM
Type of Device
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive,
suite 200
irvine CA 92618
MDR Report Key16077249
MDR Text Key306419397
Report Number16077249
Device Sequence Number1
Product Code DQK
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source User Facility
Reporter Occupation Risk Manager
Type of Report Initial
Report Date 12/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Date Report Sent to FDA12/13/2022
Event Location Hospital
Date Report to Manufacturer12/30/2022
Initial Date Manufacturer Received Not provided
Initial Date FDA Received12/30/2022
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age20440 DA
Patient SexMale
Patient Weight28 KG
Patient RaceWhite
-
-