Model Number 0673 |
Device Problems
Failure to Sense (1559); Inappropriate/Inadequate Shock/Stimulation (1574); Low impedance (2285); Device Dislodged or Dislocated (2923)
|
Patient Problem
Electric Shock (2554)
|
Event Date 12/01/2022 |
Event Type
Injury
|
Event Description
|
It was reported that this right ventricular (rv) lead was explanted due to low impedance measurements and low sensing days after implant.During lead revision, lead dislodgement to the right atrium was noticed.In addition, the patient the patient received and inappropriate shock for atrial fibrillation (af).The rv lead was explanted and replaced.No additional adverse patient effects were reported.
|
|
Manufacturer Narrative
|
Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the lead was performed.Testing was completed to assess lead electrical performance and inner/outer insulation integrity.Measurements throughout these tests were within normal limits.Microscopic inspections of the terminal pin assembly, lead body, and electrode tip found no anomalies other than blood or body fluid in the lumen due to damaged insulation and cut in the suture sleeve likely due to explant damage.Laboratory testing was unable to reproduce the reported clinical observations and detailed analysis did not reveal any abnormalities.The reported issue is issue covered by the instructions for use (ifu) - and therefore is deemed a known inherent risk of the device.
|
|
Event Description
|
It was reported that this right ventricular (rv) lead was explanted due to low impedance measurements and low sensing days after implant.During lead revision, lead dislodgement to the right atrium was noticed.In addition, the patient the patient received and inappropriate shock for atrial fibrillation (af).The rv lead was explanted and replaced.No additional adverse patient effects were reported.
|
|
Search Alerts/Recalls
|