Model Number ALLURA XPER FD20
Device Problems Failure to Power Up (1476); Activation Failure (3270)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/14/2022
Event Type  malfunction  
Event Description
It has been reported to philips that the allura system did not power on.No harm has been reported to philips.A philips service engineer inspected the system on site and identified that the fuse in the mains power distribution unit failed.The engineer replaced the 750ma fuse and the system was returned to use in good working order.
 
Manufacturer Narrative
Philips has investigated this complaint.According to the additional information, the device was in clinical use.The philips field service engineer (fse) inspected the system onsite and confirmed the system was unable to power on.Upon troubleshooting fse checked the mpd control unit and confirmed the issue occurred due to defect in the fuse.Fse replaced the fuse.After replacement of fuse, the system was returned to good working condition.The codes were updated based on the investigation outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURA XPER FD
Type of Device
SYSTEM, X-RAY, ANGIOGRAPHIC
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL  5684 PC
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL   5684 PC
Manufacturer Contact
dusty leppert
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key16078137
MDR Text Key306435054
Report Number3003768277-2022-01881
Device Sequence Number1
Product Code IZI
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K130842
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberALLURA XPER FD20
Device Catalogue Number722012
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/14/2022
Initial Date FDA Received12/30/2022
Supplement Dates Manufacturer Received07/25/2023
Supplement Dates FDA Received10/27/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/26/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-