Model Number 3100 A
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/08/2022
Event Type  malfunction  
Manufacturer Narrative
Vyaire medical file identification: (b)(4).At this time, the suspect device has not been returned for evaluation.Therefore, root cause has not been determined yet.Vyaire dispatched an fse under total service contract to complete the service.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Event Description
It was reported to vyaire medical problem with a 3100a ventilator where the bias flow meter is not working.Furthermore, there was no patient associated with the reported event.
 
Manufacturer Narrative
Device evaluation : g3, g6, h2, h3, h6 and h10.Results of investigation: vyaire medical was able to confirm the reported issue.The suspect device has not been returned for evaluation.However, vyaire reached out to customer for clarity of issue.Troubleshot issue via phone and found a defective blender.Customer was advised to replace blender and to check if issue still persist.Customer confirmed that when the blender was replaced, their problem has been resolved.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV)
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
VYAIRE MEDICAL
26125 n. riverwoods blvd.
mettawa IL 60045
Manufacturer (Section G)
VYAIRE MEDICAL INC.
1100 bird center drive
palm springs CA 92262
Manufacturer Contact
erika bonilla
510 technology drive
irvine, CA 92618
MDR Report Key16079450
MDR Text Key308476749
Report Number2021710-2022-17079
Device Sequence Number1
Product Code LSZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3100 A
Device Catalogue Number768901
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/08/2022
Initial Date FDA Received12/30/2022
Supplement Dates Manufacturer Received07/28/2023
Supplement Dates FDA Received08/01/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/26/1998
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-