Analysis and results: there are no previous complaints of this code batch of which we manufactured and distributed in the market (b)(4) units.There are no units in stock in b.Braun surgical's warehouse.We have received 2 closed samples.We have tested the knot pull tensile strength of the closed samples received and the results fulfill the requirements of the european pharmacopoeia (ep): 1.22 kgf in average and 1.21 kgf in minimum (ep requirements: 0.51 kgf in average and 0.15 kgf in minimum).We have also tested the needle penetration performance (1st penetration) of the two needles received and the results fulfill b.Braun surgical requirements: 0.479 n in average and the maximum for the 1st penetration average is 0.480 n.Reviewed the batch manufacturing record, this product had a normal process and the results during the process fulfill usp/ep and b.Braun surgical requirements.Final conclusion: although the results of the samples received fulfil the specifications of european pharmacopoeia/ b.Braun surgical specifications, we take note of this incidence in order to assess if new or additional actions are needed.The case is considered not confirmed by evidence of the samples received.Actions on product distributed of this reference/batch: based on the conclusion derived from investigation, it is not required to make actions in distributed product.Corrective/preventive actions: according to our internal procedures, there is no need to establish corrective or preventive actions.Nevertheless, this complaint is recorded for trending analysis to assess if actions are needed in the future.
|