CORDIS CORPORATION CATH TEMPO 4F VER 135 DEGREE 100CM; CATHETER, INTRAVASCULAR, DIAGNOSTIC Back to Search Results
Model Number 451414H0
Device Problem Device-Device Incompatibility (2919)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/04/2022
Event Type  malfunction  
Event Description
As reported, there was resistance when attaching the syringe to the hub of two 4f tempo vertebral 100cm diagnostic catheters.It was noticed that the syringe had resistance connecting to the hub during prep.Another unknown vertebral catheter was opened and worked fine.There was no reported patient injury.The device was inspected in the package prior to opening.The device was not resterilized.The device was opened in a sterile field.The user was trained to the device.The device was stored per labeling.Other procedural details were requested but are unknown, unavailable, not answered, or not applicable.The devices will be returned for evaluation.
 
Manufacturer Narrative
This report is related to report #: 9616099-2022-06070.A review of the manufacturing documentation associated with lot 18092072 presented no issues during the manufacturing process that can be related to the reported event.This device is available for analysis but the engineering report is not yet available.However, it will be submitted within 30 days upon receipt.Additional information is pending and will be submitted within 30 days upon receipt.
 
Manufacturer Narrative
This report is related to report # 9616099-2022-06070.As reported, there was resistance when attaching the syringe to the hub of two 4f tempo vertebral 100cm diagnostic catheters.It was noticed that the syringe had resistance connecting to the hub during prep.Another unknown vertebral catheter was opened and worked fine.There was no reported patient injury.The device was inspected in the package prior to opening.The device was not resterilized.The device was opened in a sterile field.The user was trained to the device.The device was stored per labeling.Two non-sterile units of catheter cath tempo 4f ver 135 degree 100cm were received for analysis.The units were randomly identified as unit 1 and unit 2.Unit one was evaluated under complaint (b)(4) and unit 2 was evaluated under (b)(4).(b)(4): during visual inspection, no anomalies were noted on the device (brite tip/distal tip, body/shaft nor the hub).Dimensional analysis was performed to verify the correct thread outer diameter and found within specification.An attempt to connect the unit with a lab sample syringe was successfully performed and no difficulty was noted.A flushing test was performed, and no leakage or anomaly was noted.A product history record (phr) review of lot 18092072 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.(b)(4): during visual inspection, no anomalies were noted on the device (brite tip/distal tip, body/shaft nor the hub).Dimensional analysis was performed to verify the correct thread outer diameter and found within specification.An attempt to connect the unit with a lab sample syringe was successfully performed and no difficulty was noted.A flushing test was performed, and no leakage or anomaly was noted.A product history record (phr) review of lot 18092072 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.The events reported by the customer as ¿luer hub - catheters ¿ incompatibility/fit - with syringe¿ was not confirmed for either device since they were connected to a lab sample syringe with no difficulty or leaks noted.Handling factors and/or issues with the syringe may have contributed to the reported event.Users are trained to inspect for signs of damage prior to and during use.Any product with damage is not to be used.Information for safety is provided in the products labeling with the intent to make the user aware of the risks.According to the instructions for use (ifu), although not intended as a mitigation of risk, ¿treat all 4f (1.35 mm) catheters and smaller french sizes with ultimate care.The performance of these products may be impaired if not properly and cautiously handled during unpacking and preparation.¿ based on the information available and the phr review, there is no indication that the event is related to the device design or manufacturing process.Therefore, no preventative or corrective actions will be taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CATH TEMPO 4F VER 135 DEGREE 100CM
Type of Device
CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
CORDIS CORPORATION
cordis corporation
miami lakes FL 33014
Manufacturer (Section G)
CORDIS CORPORATION
14201 nw 60 avenue
miami lakes FL 33014
Manufacturer Contact
karla castro
14201 nw 60 avenue
miami lakes, FL 33014
7863138372
MDR Report Key16084116
MDR Text Key306790876
Report Number9616099-2023-06199
Device Sequence Number1
Product Code DQO
Combination Product (y/n)Y
Reporter Country CodeAS
PMA/PMN Number
K973401
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number451414H0
Device Catalogue Number451414H0
Device Lot Number18092072
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/08/2022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/07/2022
Initial Date FDA Received01/02/2023
Supplement Dates Manufacturer Received01/06/2023
Supplement Dates FDA Received01/11/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/03/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
UNKNOWN VERTEBRAL CATHETER
-
-