EDWARDS LIFESCIENCES DR TRUWAVE DISPOSABLE PRESSURE TRANSDUCER; TRANSDUCER, PRESSURE, CATHETER TIP
|
Back to Search Results |
|
Model Number PX260 |
Device Problem
Device Contaminated During Manufacture or Shipping (2969)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/09/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
One pressure monitoring set was received by our product evaluation laboratory for a full examination.The report of foreign particle in tubing was confirmed.An unknown dark brown particulate, approximately 9.5 cm proximal from pressure line distal male connector was observed inside pressure tubing.The particulate was approximately 0.4"x 0.1" in size, attached to the inner wall of pressure tubing.The particulate stayed at the same location on pressure tubing after 5 minutes of continuous flushing.The particulate was sent to chemistry lab for further analysis and the ir spectrum of dark unknown material showed similar absorption characteristics when comparing to poly(vinyl-chloride) like material.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
|
|
Event Description
|
As reported, during flushing with this pressure monitoring set, a visible black foreign material was found inside the patient tubing, near the connector.Patient demographics unable to be obtained.There was no allegation of patient injury.The device was available for evaluation.
|
|
Manufacturer Narrative
|
The manufacturing records were reviewed for the lot involved and there is no indication of a related nonconformance.All process parameters were met and inspections passed successfully.Based on further engineering investigation it was determined that this issue is related to the supplier process; therefore, the supplier has been notified in order to perform an investigation to avoid the recurrence of this issue.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
|
|