• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. ALLURA XPER FD; SYSTEM, X-RAY, ANGIOGRAPHIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. ALLURA XPER FD; SYSTEM, X-RAY, ANGIOGRAPHIC Back to Search Results
Model Number ALLURA XPER FD20
Device Problems Positioning Failure (1158); Insufficient Information (3190)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/22/2022
Event Type  malfunction  
Event Description
It was reported to philips that the control module of the device was deformed.The device was inside clinical use at the time of the event.No patient harm was reported.To date, no further information was received.We are conservatively reporting this event as the investigation is ongoing.A follow-up report will be submitted when further information is received.
 
Manufacturer Narrative
Philips has investigated this complaint.The philips field service engineer (fse) inspected the system onsite and confirmed that the geometry module of the device was deformed.Review of the system log file showed that there was a malfunction with geo pc.To resolve the issue, fse replaced the geo module and after that, the system was returned to use in good working order.The codes were updated based on the investigation outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURA XPER FD
Type of Device
SYSTEM, X-RAY, ANGIOGRAPHIC
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL  5684 PC
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL   5684 PC
Manufacturer Contact
dusty leppert
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key16087788
MDR Text Key306544541
Report Number3003768277-2023-00010
Device Sequence Number1
Product Code IZI
UDI-Device Identifier00884838059054
UDI-Public00884838059054
Combination Product (y/n)N
Reporter Country CodeTU
PMA/PMN Number
K130842
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberALLURA XPER FD20
Device Catalogue Number722012
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/22/2022
Initial Date FDA Received01/03/2023
Supplement Dates Manufacturer Received01/17/2024
Supplement Dates FDA Received01/25/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/10/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-