Model Number CYF-VA2
Device Problem Scratched Material (3020)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
An evaluation of the returned device confirmed the customer allegation.Connecting tube had a dent.There were few additional findings.Adhesive on bending section cover was chipped.Connecting tube was wrinkled.Due to wear of angle wire, bending angle in up direction does not meet the standard value.Electrical contact of video connector was shaved.Label on light guide connector was peeled.Universal cord was wrinkled and dented.Scratches were found throughout the device.The investigation is ongoing and follow up with the user facility is currently being performed.A supplemental will be submitted on completion of investigation or if any additional information is available.
 
Event Description
The company representative on behalf of the customer reported, the image guide tube of the loaner cysto-nephro videoscope was corrugated.The device was returned, and an evaluation revealed scraping of the forceps plug cap.This report is being submitted to capture the reportable malfunction found during evaluation.There were no reports of patient harm associated with the event.
 
Event Description
It was corrected that the device was owned by the manufacturer; there was no reported complaint from users who had used this device.The suggested phenomenon was confirmed on inspection after it was returned.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 14 years since the subject device was manufactured.Based on the results of the investigation, a definitive root cause could not be established.The suggested phenomenon was presumed to have been due to stress, handling, or other factors.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISERA CYSTO-NEPHRO VIDEOSCOPE
Type of Device
CYSTO-NEPHRO VIDEOSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16094473
MDR Text Key308623675
Report Number3002808148-2023-00058
Device Sequence Number1
Product Code FAJ
UDI-Device Identifier04953170339455
UDI-Public04953170339455
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K133538
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 01/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCYF-VA2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/06/2022
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/06/2022
Initial Date FDA Received01/04/2023
Supplement Dates Manufacturer Received01/10/2023
Supplement Dates FDA Received01/30/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/15/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse