• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SPINE INC VIPER 2 SYSTEM COMPRESSION/DISTRACTION RACK 5.5; EXTRACTOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SPINE INC VIPER 2 SYSTEM COMPRESSION/DISTRACTION RACK 5.5; EXTRACTOR Back to Search Results
Model Number 286740020
Device Problem Device Slipped (1584)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/09/2022
Event Type  malfunction  
Event Description
Device report from synthes reports an event in portugal as follows: it was reported that during a procedure on an unknown date, the viper compressor/distractor rack was loose when it was positioned either for compression of distraction.Surgery was delayed by 5 minutes due to the event.The procedure was successfully completed with no adverse consequence to the patient.No further information is available.This report involves one viper 2 system compression/distraction rack 5.5.This is report 1 of 1 for (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Only the event year is known.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Enitial reporter facility address: (b)(4).Reporter is a j&j employee.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.The device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.If information is obtained that was not available for this medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h6: the investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.If information is obtained that was not available for this medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.B3: event date added.H4: manufacture date added.Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device history lot: a review of the receiving inspection (ri) for viper3d compressor/distractor was conducted identifying that lot number gm5134402 released in three batches.¿ batch1: lot qty of (b)(4) units were released on 30 oct 2018 with no discrepancies.¿ batch2: lot qty of (b)(4) units were released on 11 dec 2018 & 05 feb 2019 with no discrepancies.¿ batch3: lot qty of (b)(4) units were released on 10 oct 2019 with no discrepancies.Supplier: seabrook medical.As a result, the ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.Device history review: the ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.B5: event description updated.D10: concomitant product added.
 
Event Description
It was detected during inspection that there were two products from this surgery that were non-conforming, despite only one being reported initially.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: the product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the viper3d compressor/distractor was found the device loose from the locking mechanism.A dimensional inspection for the viper3d compressor/distractor was unable to be performed as it is not applicable to the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the viper3d compressor/distractor would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Drawing/specifications reviewed? yes, reviewed.Dimensional inspection: n/a.Device history a review of the receiving inspection (ri) for viper3d compressor/distractor was conducted identifying that lot number gm5134402 released in three batches.¿ batch1: lot qty of (b)(4) units were released on 30 oct 2018 with no discrepancies.¿ batch2: lot qty of (b)(4) units were released on 11 dec 2018 & 05 feb 2019 with no discrepancies.¿ batch3: lot qty of (b)(4) units were released on 10 oct 2019 with no discrepancies.Supplier:(b)(4) as a result, the ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIPER 2 SYSTEM COMPRESSION/DISTRACTION RACK 5.5
Type of Device
EXTRACTOR
Manufacturer (Section D)
DEPUY SPINE INC
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key16095146
MDR Text Key306625435
Report Number1526439-2023-00008
Device Sequence Number1
Product Code HWB
UDI-Device Identifier10705034214757
UDI-Public10705034214757
Combination Product (y/n)N
Reporter Country CodePO
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number286740020
Device Catalogue Number286740020
Device Lot NumberGM5134402
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/10/2022
Initial Date FDA Received01/04/2023
Supplement Dates Manufacturer Received01/16/2023
02/10/2023
03/13/2023
05/19/2023
Supplement Dates FDA Received01/25/2023
02/28/2023
03/29/2023
05/25/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/30/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
VIPER3D COMPRESSOR/DISTRACTOR
-
-