OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.2 WIRE, MEDIUM, 30°, STERILE, SINGLE USE Back to Search Results
Model Number WA22306D
Device Problem Unintended Movement (3026)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/06/2022
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress; therefore, the root cause of the reported defect cannot be determined at this time.However, if additional information becomes available, this report will be supplemented accordingly.In general, the customer is required to inspect the device for defects, check the function of all devices and have alternate equipment prior to use.Olympus will continue to monitor the field performance of this device.
 
Event Description
Olympus was informed that while using a resection electrode, in the second half of the case-use procedure, when the doctor was incising the mucous membrane with cut output, a problem occurred in which the blue and yellow at both ends of the loop gradually moved inward and the loop itself became narrower.The device was replaced with a new product and completed the procedure.No death, injury or harm has been reported to olympus.
 
Event Description
The reported event occurred during a therapeutic transurethral resection of bladder tumor procedure.
 
Manufacturer Narrative
Corrected fields: b5/d1/d4/d9 (information was inadvertently omitted from the initial), g2 (add other and remove health professional).This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and device evaluation.The device was returned to olympus for inspection, and the customer's complaint was confirmed.The loop electrode at the tip was deformed.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause could not be determined.It is likely the reported event occurred due to the use of excessive force from improper handling.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.2 WIRE, MEDIUM, 30°, STERILE, SINGLE USE
Type of Device
ELECTRODE
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16095219
MDR Text Key308619033
Report Number9610773-2023-00071
Device Sequence Number1
Product Code FAS
UDI-Device Identifier14042761051689
UDI-Public14042761051689
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 05/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberWA22306D
Device Catalogue NumberWA22306D
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/06/2022
Initial Date FDA Received01/04/2023
Supplement Dates Manufacturer Received05/11/2023
Supplement Dates FDA Received05/31/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/01/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-