Statement of the reported problem: this report is based on information provided by philips response service engineer (rse) and has been investigated by the philips complaint handling team.Philips received a complaint on the suresigns vs2+ nbp/spo2/wireless indicating that the audio failed and an inop message was displayed.There was no report of a death or serious injury, nor was there a report of any adverse impact to any user or patient.Functional testing/service repair/technical investigation: the following functional tests were performed: the rse spoke to the customer and determined that the loss of audio was the result of a failed speaker.Confirmation: based on the information available and the testing conducted, the cause of the reported problem was a failed speaker.The reported problem was confirmed.Further action decision: based on the information available and results of additional analysis, no further action is necessary at this time.Closure & product disposition: serviceable ¿ replacement product provided: the customer placed an order for a new speaker (ss mechasy speaker assy-fs part number - 453564293611) and is taking responsibility for performing the device repair to resolve the issue.It has been concluded that no further action is required at this time.If additional information is received the complaint file will be reopened.
|