SHIRAKAWA OLYMPUS CO., LTD. HIGH FLOW INSUFFLATION UNIT Back to Search Results
Model Number UHI-4
Device Problems Break (1069); Failure to Clean Adequately (4048)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
During a standard inspection of a customer-returned asset, the high flow insufflation unit presented with a leak from the first regulator unit.The customer did not report any problems associated with the device, and there was no patient/user injury or harm reported to olympus.
 
Manufacturer Narrative
Additional issues were identified during the device evaluation: low insufflation due to a defective manifold unit; the main switch had a crack; the front panel gasket was dirty; and the flat cable was corroded.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the leakage from the defective manifold unit was due to a failure of the safety valve of the primary pressure reducer.However, the definitive root causes of the safety valve failure, dirty front panel gasket, and flat cable rust could not be determined.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
HIGH FLOW INSUFFLATION UNIT
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16106207
MDR Text Key308575118
Report Number3002808148-2023-00126
Device Sequence Number1
Product Code HIF
UDI-Device Identifier04953170324147
UDI-Public04953170324147
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K122180
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUHI-4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/13/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/13/2022
Initial Date FDA Received01/05/2023
Supplement Dates Manufacturer Received03/03/2023
Supplement Dates FDA Received03/31/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/24/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-