H10: five (5) devices were received for evaluation.Visual inspection was performed and observed a primary packaging seal not completely sealed in various areas of the samples (four samples were found with incomplete top seal, one sample with an incomplete bottom side seal, and one sample with incomplete top and bottom seals).The reported condition was verified.The cause of the condition could not be determined; however, the possible causes for the seal defects found include impact damage sustained during transport or to variation during the manufacturing packaging process.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|