• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION EUROPEAN VALVE SET, EM2400; SET, I.V. FLUID TRANSFER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION EUROPEAN VALVE SET, EM2400; SET, I.V. FLUID TRANSFER Back to Search Results
Catalog Number H938724E
Device Problem Unsealed Device Packaging (1444)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that the pouch seals of five em2400 valve sets were incomplete in places.This was observed before use at the compounding facility.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
H10: five (5) devices were received for evaluation.Visual inspection was performed and observed a primary packaging seal not completely sealed in various areas of the samples (four samples were found with incomplete top seal, one sample with an incomplete bottom side seal, and one sample with incomplete top and bottom seals).The reported condition was verified.The cause of the condition could not be determined; however, the possible causes for the seal defects found include impact damage sustained during transport or to variation during the manufacturing packaging process.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EUROPEAN VALVE SET, EM2400
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
AVAILMED
c. industrial lt. 001 mz. 105
no 20905 int a, col cd ind.
tijuana, baja california 22444
MX   22444
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key16109252
MDR Text Key308388771
Report Number1416980-2022-07275
Device Sequence Number1
Product Code LHI
UDI-Device Identifier00085412475776
UDI-Public(01)00085412475776
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 02/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberH938724E
Device Lot Number60382999
Was Device Available for Evaluation? Device Returned to Manufacturer
Initial Date Manufacturer Received 12/13/2022
Initial Date FDA Received01/05/2023
Supplement Dates Manufacturer Received01/26/2023
Supplement Dates FDA Received02/07/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/20/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-