• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL MASTERS SERIES HEART VALSALVA AORTIC VALVES GRAFT; HEART-VALVE, MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL MASTERS SERIES HEART VALSALVA AORTIC VALVES GRAFT; HEART-VALVE, MECHANICAL Back to Search Results
Model Number 25VAVGJ-515
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Hematoma (1884); Pleural Effusion (2010); Tachycardia (2095); Cardiac Tamponade (2226)
Event Date 12/12/2022
Event Type  Injury  
Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event Description
(b)(4) - valved grafts pas, patient site id: (b)(6).It was reported that on (b)(6) 2022, a 25mm sjm masters series valsalva aortic valved graft was implanted into a patient.It was reported that on (b)(6) 2022, the patient experienced deterioration of well-being, malaise, and tachycardia.A chest x-ray and transthoracic echocardiogram (tte) were done, showing features of a hematoma in the mediastinum between the left and right coronary artery.A resternotomy was performed due to tamponade, and a blood transfusion was given.Warfarin was prescribed.Pericardial tamponade declared resolved.It is believed that the pericardial effusion is not related to the device.The patient is reported to be discharged.No additional information was provided.
 
Manufacturer Narrative
An event of deterioration of well-being, malaise, tachycardia, a hematoma in the mediastinum between the left and right coronary artery and pericardial tamponade was reported.A returned device assessment could not be performed as the device remains implanted and was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Information from the field indicated that the reported event was not related to the device.There was no allegation of malfunction against the abbott device.Based on the information received, the cause of the reported incident could not be conclusively determined.There is no indication of a product quality issue with regards to manufacture, design, or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MASTERS SERIES HEART VALSALVA AORTIC VALVES GRAFT
Type of Device
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL PUERTO RICO, INC. REG#2648612
20 b st caguas west park
caguas 00725
*   00725
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key16117579
MDR Text Key306877867
Report Number2135147-2023-00090
Device Sequence Number1
Product Code LWQ
UDI-Device Identifier05414734009492
UDI-Public05414734009492
Combination Product (y/n)N
Reporter Country CodePL
PMA/PMN Number
P810002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Study,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number25VAVGJ-515
Device Catalogue Number25VAVGJ-515
Device Lot Number8617582
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/14/2022
Initial Date FDA Received01/06/2023
Supplement Dates Manufacturer Received03/14/2023
Supplement Dates FDA Received03/16/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/15/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-