Model Number OER-4
Device Problem Flushing Problem (1252)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
The customer reported to olympus, the endoscope reprocessor had an e99 error and a white material attached to the scopes that were washed.The issue was found during reprocessing.There were no reports of patient harm.Related patient identifiers: (b)(6).
 
Manufacturer Narrative
The device was not returned to olympus for evaluation.The customer stated that the concentration was not checked every time, instead it was performed after 2 weeks.A supplemental report will be submitted if any additional information is provided by the user facility.
 
Manufacturer Narrative
This supplemental report is being submitted to indicate that this is a duplicate complaint of (b)(4), which was submitted under 9610595 - 2023 - 00361.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOSCOPE REPROCESSOR
Type of Device
ENDOSCOPE REPROCESSOR
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16118188
MDR Text Key308830476
Report Number9610595-2023-00358
Device Sequence Number1
Product Code FEB
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 01/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOER-4
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/09/2022
Initial Date FDA Received01/06/2023
Supplement Dates Manufacturer Received01/11/2023
Supplement Dates FDA Received01/11/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/13/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-