This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the definitive root cause of the foreign material and concentration check issue could not be determined.It is possible that failure of secured precleaning before the manual cleaning may have left some kind of uncleaned material remained and that the user mishandled the management method of the disinfectant solution.The instruction manual identifies the following verbiage, which may have prevented the phenomenon: ¿chapter 4 basic endoscope reprocessing operations.4.3 endoscope precleaning and manual cleaning.·perform at least the processes from precleaning to manual cleaning that involves cleaning the outer surfaces of the endoscope, brushing the area around the forceps and the suction channel, and cleaning the valves.·always preclean each endoscope immediately after the examination.If precleaning is not executed promptly, debris will solidify and may prevent effective reprocessing.·failure to preclean will leave excessive amounts of debris adhering to the endoscope and may compromise the effectiveness of the reprocessing.It may also result in debris accumulating in the endoscope, preventing the endoscope from working correctly.¿, as well as "¿chapter 3 inspection before use 3.9 inspecting the disinfectant solution¿s concentration level on reprocessing an endoscope, always use an acecide checker or a portable concentration checker to check that the disinfectant solution is at an effective concentration.Please be sure to replace the disinfectant solution before the disinfectant effect disappears.¿ olympus will continue to monitor field performance for this device.
|