• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. ALLURA XPER FD; SYSTEM, X-RAY, ANGIOGRAPHIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. ALLURA XPER FD; SYSTEM, X-RAY, ANGIOGRAPHIC Back to Search Results
Model Number ALLURA XPER FD10
Device Problem No Device Output (1435)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/06/2022
Event Type  malfunction  
Event Description
It was reported to philips that the emergency stop was activated and could not be dis-armed.
 
Manufacturer Narrative
Philips has investigated this complaint.According to additional information received, it was unknown if the issue was identified while the system was in clinical use, but no patient or user harm was reported to philips.A philips remote service engineer (rse) inspected the system remotely and confirmed the issue.Troubleshooting by the rse and log file analysis confirmed an issue with the emergency stop button on the system's geometry module.As part of the troubleshooting, the rse advised the customer to ensure that the geometry module did not have any damaged, broken, or loose cabling or connectors.After troubleshooting, the issue did not reoccur, and the system was returned to use in good working order.The codes were updated based on the investigation outcome.Evaluation method codes were corrected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURA XPER FD
Type of Device
SYSTEM, X-RAY, ANGIOGRAPHIC
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL  5684 PC
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL   5684 PC
Manufacturer Contact
dusty leppert
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key16120805
MDR Text Key306936286
Report Number3003768277-2023-00119
Device Sequence Number1
Product Code IZI
UDI-Device Identifier00884838059030
UDI-Public00884838059030
Combination Product (y/n)N
Reporter Country CodeIS
PMA/PMN Number
K130842
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberALLURA XPER FD10
Device Catalogue Number722010
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/06/2022
Initial Date FDA Received01/09/2023
Supplement Dates Manufacturer Received12/06/2022
Supplement Dates FDA Received02/07/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/28/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-