Model Number 350P
Device Problem Key or Button Unresponsive/not Working (4063)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/13/2022
Event Type  malfunction  
Manufacturer Narrative
Heartsine has requested return of the device for investigation.Upon completion, the conclusions will be submitted in a follow-up report.
 
Event Description
A distributor contacted heartsine to report that a customer¿s device was not responding to on/off button presses.In this state the device may not be able to deliver defibrillation therapy if needed.There was no report of patient use associated to the reported event.
 
Manufacturer Narrative
Heartsine has made multiple attempts to contact the customer regarding the status of their device, but no response has been received from the customer.The device was not returned to heartsine for investigation.No additional information will be forthcoming at this time.The cause of the reported issue could not be determined.If the device is returned to the manufacturer the investigation will be reopened.
 
Event Description
A distributor contacted heartsine to report that a customer¿s device was not responding to on/off button presses.In this state the device may not be able to deliver defibrillation therapy if needed.There was no report of patient use associated to the reported event.
 
Event Description
A distributor contacted heartsine to report that a customer¿s device was not responding to on/off button presses.In this state the device may not be able to deliver defibrillation therapy if needed.There was no report of patient use associated to the reported event.
 
Manufacturer Narrative
Heartsine's investigation of the device confirmed the reported fault as upon receipt the device was found to be powering on automatically.This fault was attributed to corrosion on the membrane tail.The corrosion on the membrane tail would suggest that the device had been stored outside the indicated conditions.This was further confirmed by corrosion on the screws.The device was scrapped by heartsine and the customer was provided with a replacement device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PACKAGE,350P,PP03,INTL EN,350-BAS-UK-10
Type of Device
AUTOMATED EXTERNAL DEFIBRILLATOR
Manufacturer (Section D)
HEARTSINE TECHNOLOGIES LTD
207 airport road west
belfast BT3 9 ED
EI  BT3 9ED
Manufacturer (Section G)
HEARTSINE TECHNOLOGIES LTD
207 airport road west
belfast BT3 9 ED
EI   BT3 9ED
Manufacturer Contact
todd bandy
207 airport road west
belfast BT3 9-ED
EI   BT3 9ED
4258674577
MDR Report Key16127484
MDR Text Key308562819
Report Number3004123209-2023-00003
Device Sequence Number1
Product Code NSA
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
P160008
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date 05/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number350P
Device Catalogue Number350-BAS-UK-10
Was Device Available for Evaluation? Device Returned to Manufacturer
Initial Date Manufacturer Received 12/13/2022
Initial Date FDA Received01/09/2023
Supplement Dates Manufacturer Received01/24/2023
05/09/2023
Supplement Dates FDA Received02/21/2023
05/31/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/25/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-