MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. OES CHOLEDOCHO FIBERSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. OES CHOLEDOCHO FIBERSCOPE Back to Search Results
Model Number CHF-P60
Device Problem Peeled/Delaminated (1454)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/21/2022
Event Type  malfunction  
Manufacturer Narrative
The device was returned, and an initial evaluation was conducted by olympus; however, investigation is ongoing.During the initial evaluation, the user¿s report was confirmed.The benign cover was found leaking.Additionally, as noted , the coating material on the insertion section was peeling off.There were also more than 10 black dots noted in the imaging.If additional information becomes available following the device evaluation, a supplemental report will be filed.
 
Event Description
The customer reported that while performing routine maintenance on his olympus oes choledocho fiberscope, it was discovered that the device was leaking.There was no patient involvement during this event.During the device evaluation, it was discovered that the coating material was peeling off the unit.This report is being submitted to capture the peeling coating found during the device evaluation.
 
Manufacturer Narrative
Correction: e2 was inadvertently selected on the initial report.This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the event was unable to be identified; however, it was confirmed that the coating on the insertion part had peeled off.The event likely occurred due to physical stress/chemical stress/storage environment, etc.Are presumed to be the cause.The event can be detected/prevented by following the instructions for use which state: 1) "inspect the external surface of the entire insertion tube including the bending section and the distal end for dents, bulges, swelling, peeling, scratching, holes, sagging, transformation, bends, adhesion of foreign body, dropout of parts, any protruding objects or other irregularities." 2) "warning: do not strike, bend, hit, pull, twist, or drop the endoscope¿s distal end, insertion tube, bending section, control section, universal cord, or endoscope connector with excessive force.The endoscope may be damaged and could cause patient injury, such as burns, bleeding, and/or perforation.It could also cause parts of the endoscope to fall off inside the patient." 3) "this instrument is not durable, or does not have sufficient durability against the respective methods stated in the guidelines of each country for destroying or inactivating prions.For information on the durability against each method, please contact olympus.If cleaning, disinfection and sterilization methods not stated in this instruction manual are performed, olympus cannot guarantee the effectiveness, safety and durability of this instrument.Make sure to confirm that there is no abnormality before use, and use under responsibility of a physician.Do not use if any abnormality is found." 4) "warning: the storage cabinet must be clean, dry, well ventilated and maintained at ambient temperature.Storing the endoscope in direct sunlight, at high temperatures, in high humidity or exposed to ozone, x-rays and/or ultraviolet-rays may damage the endoscope or present an infection control risk." olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OES CHOLEDOCHO FIBERSCOPE
Type of Device
CHOLEDOCHO FIBERSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16130664
MDR Text Key308869836
Report Number9610595-2023-00487
Device Sequence Number1
Product Code FBO
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K843084
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCHF-P60
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/23/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/21/2022
Initial Date FDA Received01/10/2023
Supplement Dates Manufacturer Received01/31/2023
Supplement Dates FDA Received02/20/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/20/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-