ELEKTA SOLUTIONS AB MOSAIQ; ACCELERATOR, LINEAR, MEDICAL Back to Search Results
Model Number MOSAIQ
Device Problem Computer Software Problem (1112)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/12/2022
Event Type  malfunction  
Event Description
The customer reported intermittent mosaiq error messages on linac sequencer.
 
Manufacturer Narrative
The manufacturer's investigation is on-going and further information will be provided once the investigation has completed.
 
Manufacturer Narrative
The investigation was completed by conducting a thorough evaluation of the product and the reported information.The customer reported intermittent mosaiq error messages on linac sequencer.It was ascertained from the mosaiq logs that the cause of the clientvmi and com call failed messages encountered by the customer as well as the patient mismatch whilst selecting the next patient appears to be the result of communication/network issues which produced an inconsistent state.Mosaiq did not have any malfunction and is working as designed and intended.Based on the available information there has been no mistreatment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOSAIQ
Type of Device
ACCELERATOR, LINEAR, MEDICAL
Manufacturer (Section D)
ELEKTA SOLUTIONS AB
kungstensgatan 18
box 7593
stockholm, SE-10 3 93
SW  SE-103 93
Manufacturer (Section G)
ELEKTA SOLUTIONS AB
400 perimeter center terrace
suite 50
atlanta GA 30346
Manufacturer Contact
pms
cornerstone
london road
crawley, west sussex RH10 -9BL
UK   RH10 9BL
MDR Report Key16131441
MDR Text Key307511362
Report Number3015232217-2023-00000
Device Sequence Number1
Product Code IYE
UDI-Device Identifier00858164002367
UDI-Public(01)00858164002367(10)2.83.40
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K183034
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 08/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberMOSAIQ
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 12/12/2022
Initial Date FDA Received01/10/2023
Supplement Dates Manufacturer Received12/12/2022
Supplement Dates FDA Received08/02/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/30/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-