• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION SPACEOAR SYSTEM; ABSORBABLE PERIRECTAL SPACER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION SPACEOAR SYSTEM; ABSORBABLE PERIRECTAL SPACER Back to Search Results
Device Problems Fluid/Blood Leak (1250); Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/15/2022
Event Type  malfunction  
Manufacturer Narrative
The complainant was unable to report the upn and lot number; therefore, the manufacture date and expiration date are unknown.(b)(4).
 
Event Description
It was reported to boston scientific corporation that during a spaceoar implant procedure performed on (b)(6) 2022, fluid was leaking between the accelerator syringe and y-connector.The procedure was performed under general anesthesia and fiducials were administered transperineally.However, the procedure was ultimately aborted due to physician discretion.There were no patient complications reported as result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPACEOAR SYSTEM
Type of Device
ABSORBABLE PERIRECTAL SPACER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
500 commander shea boulevard
quincy MA 02171
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key16132640
MDR Text Key308003800
Report Number3005099803-2023-00014
Device Sequence Number1
Product Code OVB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K181465
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 01/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/15/2022
Initial Date FDA Received01/10/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient SexMale
-
-