Catalog Number 8065751155
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/01/2022
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A physician reported balanced salt solution (bss) bottle air supply tube of the pak was disconnected during surgery.It was reinserted and the surgery was continued.The product was replaced and the surgery was completed.There was no patient harm.
 
Manufacturer Narrative
A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.The returned sample was visually inspected.Visual inspection confirmed gray tube was detached from drip chamber.Insertion mark and glue-like foreign material was observed on the tube.The root cause of the customer¿s complaint is an error that occurred during the supplier¿s assembly process.Based on the condition of the open gap, it appeared the gray airline was not fully inserted into the drip chamber spike.Action will not be taken based on this occurrence.The supplier has been made aware of the issue.No adverse trends have been observed associated with the reported product and event.Quality assurance will continue to monitor and will take action for future occurrences as deemed necessary.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION SURGICAL PROCEDURE PACK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key16135370
MDR Text Key307075078
Report Number1644019-2023-00042
Device Sequence Number1
Product Code LRO
UDI-Device Identifier00380657511556
UDI-Public00380657511556
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2024
Device Catalogue Number8065751155
Device Lot Number149RLH
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/14/2022
Initial Date FDA Received01/11/2023
Supplement Dates Manufacturer Received05/16/2023
Supplement Dates FDA Received06/12/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/11/2022
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CONSTELLATION VISION SYSTEM
-
-